Body fluid analysis by 'Automated' Sysmex XN 1000 – The way ahead
DOI:
https://doi.org/10.56501/intjclinicopatholcorrel.v8i1.1075Keywords:
Sysmex XN 1000, Body fluid analysis, Cerebrospinal fluid analysis, Improved Neubauer ChamberAbstract
Body fluids are those excreted or secreted from the body. In the laboratory, the use of the term “body fluids” designates a category that excludes blood and urine. Determining the etiologic cause of fluid accumulation in various body cavities (i.e., joints, chest, and abdomen) is critical for proper treatment of various disorders. To evaluate the correlation and agreement of leucocyte and erythrocyte count in body fluids among Improved Neubauer Chamber and automated method using Sysmex XN-1000 body fluid module. A total of 500 routinely collected body fluid samples (CSF, ascitic fluid, pericardial fluid and pleural fluid) were included in the study. Total WBC counts were performed manually using improved Neubauer chamber and microscopic differential counts were performed using Leishman stained slides. Corresponding results of Sysmex XN-1000 body fluid module analysis were matched to these particular count categories. Absolute WBC cell counts showed a high correlation and agreement between methods (r> 0.9, p value=<0.001). However, a fair correlation existed between the lymphocyte percentage and neutrophil percentage measured through the two methods (r= 0.436 and r=0.447 respectively, p < 0.001). Body fluid module application of Sysmex XN-1000 serves as an acceptable alternative to manual method for measuring the total nucleated cell counts and for total RBC count. However, it cannot be recommended as a suitable alternative for manual differential cytologic workup.
Keywords: Sysmex XN 1000, Body fluid analysis, Cerebrospinal fluid analysis, Improved Neubauer Chamber
Publication Facts
Reviewer profiles N/A
Author statements
Indexed in
-
—
- Academic society
- N/A
- Publisher
- MM Publications
References
Kjeldsberg CL, Knight JA. Cerebrospinal fluid. In: Body Fluids. 3rd ed. Chicago, IL: ASCP Press; 1993:65-157.
Piraino B, Bailie GR, Bernardini J, Boeschoten E, Gupta A, Holmes C, et al. Peritoneal dialysis-related infections recommendations: 2005 update. Perit Dial Int.2005;25:107–31.
Linssen J, Jennissen V, Hildmann J, Reisinger E, Schindler J, Malchau G, et al. Identification and quantification of high fluorescence stained lymphocytes as antibodies synthesizing/ secreting cells using the automated routine hematology analyser XE-2100. Cytometry B Clin Cytom 2007;72:157-66.
https://doi.org/10.1002/cyto.b.20150.
Paris A, Nhan T, Cornet E, Perol JP, Malet M, Troussard X . Performance evaluation of the body fluid mode on the platform Sysmex XE-5000 series automated hematology analyzer. Int J Lab Hematol. 2010;32(5):539–547.
https://doi.org/10.1111/j.1751-553X.2010.01220.x
Nagisa Nakazawa, Noriko Wada, Yuzo Tanaka, Kazumi Gondo, Yumiko Tanaka, Takayuki Seto, et al. Evaluation of Assay Performance of the Body Fluid Mode on the Automated Hematology Analyzer XN-2000.Sysmex J Int. 2012;22(1):1-11.
C. Fleming, R. Brouwer, J. Lindemans, R. de Jonge. Validation of the body fluid module on the new Sysmex XN-1000 for counting blood cells in cerebrospinal fluid and other body fluids. Clin Chem Lab Med 2013;50(10):1791-1798. https://doi.org/10.1515/cclm-2011-0927
Lehto T M, Leskinen P, Hedberg P and Vaskivuo T. Evaluation of the Sysmex XT-4000i for the automated body fluid analysis. Int J Lab Hematol.2014;36(2):114-123.https://doi.org/10.1111/ijlh.12134
Hoffmann JJ, Janssen WC. Automated counting of cells in cerebrospinal fluid using the CellDyn-4000 hematology analyzer. Clin Chem Lab Med 2002; 40:1168–73.
https://doi.org/10.1515/CCLM.2002.204
Aulesa C, Mainar I, Prieto M, Cobos N, Galimany R. Use of the Advia 120 hematology analyzer in the differential cytologic analysis of biological fluids (cerebrospinal, peritoneal, pleural, pericardial, synovial, and others). Lab Hematol. 2003;9(4):214-24
Andrea Perné et al. Performance Evaluation of the Sysmex XE-5000 Hematology Analyzer for White Blood Cell Analysis in Cerebrospinal Fluid. Archives of Pathology & Laboratory Medicine.2012;136(2): 194- 198.
https://doi.org/10.5858/arpa.2011-0030-OA
de Jonge R, Brouwer R, van Rijn M, van Acker BA, Otten HJ, Lindemans J. Automated analysis of pleural fluid total and differential leukocyte counts with the Sysmex XE-2100. Clin Chem Lab Med 2006;44:1367–71.
https://doi.org/10.1515/CCLM.2006.242
Conner BD, Lee YC, Branca P, Rogers JT, Rodriguez RM,Light RW. Variations in pleural fluid WBC count and differential counts with different sample containers and different methods. Chest 2003; 123:1181–7.
https://doi.org/10.1515/CCLM.2006.242
James E. Williams, Jeri Walters, Kimberley Kabb, Gaining Efficiency in the Laboratory – Automated Body Fluid Cell Counts: Evaluation of the Body Fluid Application on the Sysmex XE-5000 Hematology Analyzer. Lab Medicine.2011; 42(7):395-401.
https://doi.org/10.1309/V1HMLAYJRY01RT
Busadee Pratumvinit, Noppawan Sivasariyanonds, Leatchai Wachirutmanggur, Wimol Chinsawangwatanakul, Sathien Sukpanichnant. Comparison of Conventional Manual Methods with the ADVIA 120 Automated Method for Counting of Red and White Blood Cells in Cerebrospinal Fluid. Siriraj Med J 2007; 59: 64-68.
Takemura H, Tabe Y, Ishii K , et al. Evaluation of capability of cell count and detection of tumor cells in cerebrospinal and body fluids by automated hematology analyzer [in Japanese]. Rinsho Byori. 2010;58(6):559–564.
Boer K, Deufel T, Reinhoefer M. Evaluation of the XE-5000 for the automated analysis of blood cells in cerebrospinal fluid. Clin Biochem. 2009;42:684–691.
Aune MW, Becker JL, Brugnara C, Canfield W, Dorfman DM,Fiehn W, et al. Automated flow cytometric analysis of bloodcells in cerebrospinal fluid: analytic performance. Am J ClinPathol 2004;121:690–700.
https://doi.org/10.1016/j.clinbiochem.2009.01.025
de Jonge R, Brouwer R, van Rijn M, van Acker BA, Otten HJ, Lindemans J. Automated analysis of pleural fluid total and differential leukocyte counts with the Sysmex XE-2100. Clin Chem Lab Med 2006;44:1367–71.
https://doi.org/10.1515/CCLM.2006.242
Giuseppe Lippi, Clarissa Cattabiani1, Anna Benegiamo, Daniela Gennari, Fernanda Pavesi, Alberta Caleffi et al. Evaluation of the Fully Automated Hematological Analyzer Sysmex XE-5000 for Flow Cytometric Analysis of Peritoneal Fluid. Journal of Laboratory Automation.2013:1-5.https://doi.org/10.1177/2211068212471833
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2024 Shweta Chaturvedi, Gajendra Nath Gupta, Mansi Faujdar, Rohit Jain, Rateesh Sareen
This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.